• deutsch
  • Standard-compliant testing of side rail spacing
    under mechanical stress!

    Required as part of the safety inspection (STK) in accordance with

    EN 60601-2-52 : 2016-04 and EN 60601-2-38 : 2001-01

    201.9.1.101 *Protection against the patient becoming trapped in non-moving parts

    Checking the distances between the spars under load

    Same load test here between the rail and the lying surface

    Faster, safer, and clearer testing process

    Switch on test cone, position between side rails, pull, positive test result indicated by acoustic signal

    Test time per bed side 10-20 seconds.


    functional description

    The switch position I or II is used to select the test force (50N/250N) and to start up the device. The test cone is then positioned at the point to be tested and, by pulling on the handle, the load is applied to the grid. When the selected test force is reached, an acoustic signal is emitted, indicating that the test has been passed. If the test cone could be pulled through the side bars without a signal sounding, the test is considered failed.

    product features

    Start-up and selection of test force via switch

    Acoustic signal when load is reached

    Protection against overload via acoustic signal

    Status and condition signaling via multicolor LED

    Functional reliability thanks to tamper-proof electronics

    Usage-based service intervals

    Battery operation 9V block Voltage monitored

    Resistant to commercially available cleaning agents and disinfectants


    EBPK Electronic test cone for side rail testing

    Test cone with transport box and test gauge for distance testing

    If you have any questions about the product or would like a quote, please send your contact details to info@dibucare.de.
    The current product information is available for download here * Product information EBPK EN


    Requirements for side rails according to standard

    By definition, openings and areas must not be larger than 120 mm and must withstand testing with a defined test cone applying a test force of 250 N [EN 60601-2-52] and 50 N [EN 60601-2-38]. In section 201.9.1.101 of the standards, the corresponding openings or areas are marked with A(x). Table 201.101 describes how and at which points on the side guard (side protection) the test is to be carried out.

    Conducting the audit

    This test is considered passed if the cone could not be pulled through the opening or area being tested. The test must be carried out at the weakest point, or at several points on the side guard (side protection), if necessary. As part of the safety inspection (STK) of a nursing bed or beds generally used for medical purposes, a clear assessment of the safety of the side rails (side guards) can only be reliably determined using this defined test procedure.

    The necessity of testing side guards

    The side rails (side guards) on medical beds, which also include nursing beds, serve to protect the person in the bed. An agitated person may try to squeeze themselves between the bars of the side rails (side support) or between the bars and the lying surface with their arms, legs, head, or upper body. The bending of the spars under load can cause body parts to become trapped. This can result in bruising or even strangulation.